MPE Imaging in Tissue
MPE fluorescence microscopy is an ideal technique for
imaging tissues or other thick,
scattering specimens. Because out-of-focus
fluorescence is essentially non-existent and all photons emitted are
useful for image formation, the collection optics can be
extremely simple and therefore extremely efficient. When using single
photon confocal microscopy in tissues, a fraction of the photons emitted
in the focal volume are scattered into trajectories that do not pass
through the detector aperture,
and a fraction of the out-of-focus photons are scattered into the
aperture resulting in a blurred image. The use of MPE with widefield
collection minimizes both of these problems (1) by utilizing a
short and efficient collection path, and (2) by not producing any
out-of-focus fluorescence photons. The figure above demonstrates
this difference. A fluorescent polymer-filled EM grid (7 µm
squares) was imaged through 200 µm of water (A, confocal;
B, MPE) and through 200 µm of human dermal tissue (C, confocal;
D, MPE). Although D is degraded relative to B, the grid pattern
can still be seen. Using single photon excitation with confocal
detection on the same sample, the grid is barely recognizable (i.e.,
C compared to A).
Autofluorescence Imaging.
In addition to conventional fluorescent dyes (sections I, II, III),
MPE is useful for imaging intrinsic fluorophores as well. Biological
molecules such as NADH, elastin, collagen and the various indoleamines
are autofluorescent and MPE microscopy is the most efficient
method
for imaging their distribution in cells and tissues. The figure at right
is an MPE autofluorescence image of the elastin and collagen matrix found
in the dermal layer of skin (ex: 740 nm; em: 400-550 nm). The image was
taken ~150 µm into an slab of unfixed human dermis. Using MPE,
the extracellular structural matrices in tissues are readily visualized
as illustrated in the next section. Intercellular
components, such as indoleamines and NADH can
also be imaged.